| Treatment
Time Calculation Table - Klas series |
1) Energy loss is postulated to be linear
at 7.5% per mm of tissue.
2) Treatment time is calculated by Monte Carlo Simulation method.
| |
Treatment
Time (second) = |
Dose (J/cm3)
x 1000 x Laser Spot Size (cm²) |
| ————————————————————————— |
| End Power (mW)
x Energy Loss Factor (/cm) |
|
| Model No.: Klas-A |
 |
| Laser beam spot size: 0.13 cm2 |
|
Dose
(J/cm3) |
Depth of Tissue
(cm) |
Energy Loss Factor |
Treatment Time
(second) |
| Wavelength |
660nm |
780nm |
808nm |
| Model No. |
Klas-A50 |
Klas-A61 |
Klas-A79 |
Klas-A81 |
Klas-P62 |
Klas-P64 |
LD Power
(mW) |
50 |
100 |
90 |
110 |
260 |
460 |
End Power
(mW) |
45 |
90 |
86 |
88 |
208 |
368 |
Power Density
(mW/cm2) |
346 |
692 |
658 |
1,046 |
1,600 |
2,831 |
| 3 |
0.5 |
0.6772 |
|
26 |
13 |
13 |
13 |
6 |
3 |
| 1.0 |
0.4586 |
|
38 |
19 |
20 |
19 |
8 |
5 |
| 1.5 |
0.3105 |
|
56 |
28 |
29 |
29 |
12 |
7 |
| 2.0 |
0.2103 |
|
82 |
41 |
43 |
42 |
18 |
10 |
| 2.5 |
0.1424 |
|
|
|
64 |
62 |
26 |
15 |
| 3.0 |
0.0964 |
|
|
|
94 |
92 |
39 |
22 |
| 3.5 |
0.0653 |
|
|
|
139 |
136 |
57 |
32 |
| 4.0 |
0.0442 |
|
|
|
205 |
200 |
85 |
48 |
| 4.5 |
0.0299 |
|
|
|
303 |
296 |
125 |
71 |
| 5.0 |
0.0203 |
|
|
|
|
|
185 |
105 |
| 6.0 |
0.0093 |
|
|
|
|
|
|
228 |
|
| Model No.: Klas-E |
 |
| Laser beam spot size:
0.07 cm2 |
|
Dose
(J/cm3) |
Depth of Tissue
(cm) |
Energy Loss Factor |
Treatment Time
(second) |
| Wavelength |
660nm |
808nm |
| Model No. |
Klas-E50 |
Klas-E61 |
Klas-E81 |
Klas-P62 |
Klas-P64 |
LD Power
(mW) |
50 |
100 |
110 |
260 |
460 |
End Power
(mW) |
40 |
80 |
90 |
180 |
320 |
Power Density
(mW/cm2) |
429 |
857 |
1,100 |
2,600 |
4,600 |
| 3 |
0.5 |
0.6772 |
|
8 |
4 |
3 |
2 |
1 |
| 1.0 |
0.4586 |
|
11 |
6 |
5 |
3 |
1 |
| 1.5 |
0.3105 |
|
17 |
8 |
8 |
4 |
2 |
| 2.0 |
0.2103 |
|
25 |
12 |
11 |
6 |
3 |
| 2.5 |
0.1424 |
|
|
|
16 |
8 |
5 |
| 3.0 |
0.0964 |
|
|
|
24 |
12 |
7 |
| 3.5 |
0.0653 |
|
|
|
36 |
18 |
10 |
| 4.0 |
0.0442 |
|
|
|
53 |
26 |
15 |
| 4.5 |
0.0299 |
|
|
|
78 |
39 |
22 |
| 5.0 |
0.0203 |
|
|
|
115 |
58 |
32 |
| 6.0 |
0.0093 |
|
|
|
251 |
125 |
71 |
|
| Model No.: Klas-D |
 |
| Laser beam spot size: 0.5 cm2 |
|
Dose
(J/cm3) |
Depth of Tissue
(cm) |
Energy Loss Factor |
Treatment Time
(second) |
| Wavelength |
660nm |
808nm |
| Model No. |
Klas-D50 |
Klas-D61 |
Klas-D81 |
Klas-P62 |
Klas-P64 |
LD Power
(mW) |
50 |
100 |
110 |
260 |
460 |
End Power
(mW) |
40 |
80 |
90 |
200 |
370 |
Power Density
(mW/cm2) |
80 |
160 |
180 |
400 |
740 |
| 1 |
0.5 |
0.6772 |
|
18 |
9 |
8 |
4 |
2 |
| 1.0 |
0.4586 |
|
27 |
14 |
12 |
5 |
3 |
| 1.5 |
0.3105 |
|
40 |
20 |
18 |
8 |
4 |
| 2.0 |
0.2103 |
|
59 |
30 |
26 |
12 |
6 |
| 2.5 |
0.1424 |
|
|
|
39 |
18 |
9 |
| 3.0 |
0.0964 |
|
|
|
58 |
26 |
14 |
| 3.5 |
0.0653 |
|
|
|
85 |
38 |
21 |
| 4.0 |
0.0442 |
|
|
|
126 |
57 |
31 |
| 4.5 |
0.0299 |
|
|
|
186 |
83 |
45 |
| 5.0 |
0.0203 |
|
|
|
274 |
123 |
67 |
| 6.0 |
0.0093 |
|
|
|
|
269 |
145 |
|
| Model No.: Klas-S |
 |
| Laser beam spot size: 3 cm2 |
|
Dose
(J/cm3) |
Depth of Tissue
(cm) |
Energy Loss Factor |
Treatment Time
(minute) |
| Wavelength |
660nm |
808nm |
| Model No. |
Klas-S50 |
Klas-S61 |
Klas-S81 |
Klas-P62 |
Klas-P64 |
LD Power
(mW) |
50 |
100 |
110 |
260 |
460 |
End Power
(mW) |
30 |
60 |
44 |
104 |
184 |
Power Density
(mW/cm2) |
10 |
20 |
15 |
35 |
61 |
| 3 J/cm2 |
Surface |
1 |
|
5 |
3 |
3 |
1 |
1 |
| 3 |
0.5 |
0.6772 |
|
7 |
4 |
5 |
2 |
1 |
| 1.0 |
0.4586 |
|
11 |
5 |
7 |
3 |
2 |
| 1.5 |
0.3105 |
|
16 |
8 |
11 |
5 |
3 |
| 2.0 |
0.2103 |
|
|
12 |
16 |
7 |
4 |
| 2.5 |
0.1424 |
|
|
|
24 |
10 |
6 |
| 3.0 |
0.0964 |
|
|
|
35 |
15 |
8 |
| 3.5 |
0.0653 |
|
|
|
|
22 |
12 |
| 4.0 |
0.0442 |
|
|
|
|
33 |
18 |
| 4.5 |
0.0299 |
|
|
|
|
|
27 |
| 5.0 |
0.0203 |
|
|
|
|
|
40 |
|